Determining the Site of Modification … Part 2

September 1, 2009
by Arvin Moser, Team Manager, Application Scientists, ACD/Labs

Peak matching involves the process of comparing spectral data from a parent or starting material to an unknown compound. (The unknown compound can be referred more specifically as the product, impurity, degradant, metabolite, etc.). The similarities between the data indicate regions that have not changed while the differences indicate regions of change.

The full scan MS data in Part 1 does not offer enough spectral information to eliminate one of the candidate structures. The next step is to examine MS/MS data.

The ESI+ product ion spectra (MS/MS) for both the Parent (m/z 226) and Metabolite (m/z 228) are shown below. The spectra share a common fragment at m/z 121 and differ in the fragments at m/z 134/136 and 149/151.

MS2CompareToAPI_Part2MS2_Sept12009

Based on the fragments shown below, the suspected site of hydrogenation is on the carbonyl for metabolite A.

MS2CompareToAPI_Part2MS2Frag_Sept12009